EKG Acquisition第 43 / 100 题

A 12-lead tracing shows an inverted P wave, inverted QRS, and inverted T wave in lead I, while the chest leads look normal. What is the most likely cause?

a.The right arm and left arm electrodes are reversed
b.The V1 and V2 electrodes are reversed
c.The right leg electrode is loose
d.The paper speed is set to 50 mm/sec

解析

Reversing the right and left arm electrodes flips the polarity of lead I, producing a global inversion in that lead while the chest leads, which do not depend on arm placement, remain normal. Reversing V1 and V2 distorts only those chest leads. A loose right leg (ground) electrode causes baseline wander or noise rather than inversion, and a doubled paper speed stretches the complexes without inverting them.

免费刷完整 100 道题库 — 无需注册。

同考点相关题目

最近核对: · 审核流程

PrepPass 编辑团队 · 依据官方资料核对 NHA Certified EKG Technician (CET) Exam · 我们如何核对
反馈