Decontamination第 21 / 100 题
Instrument sets are consistently arriving in decontamination with hard, dried blood in box locks and serrations. What is the most likely root cause?
a.Point-of-use treatment is not being performed and the sets sit too long before processing
b.The washer-disinfector detergent concentration is set too high
c.The decontamination area humidity is above 60 percent
d.The transport carts are being covered too tightly
解析
Dried, baked-on soil is the classic signature of missing point-of-use treatment combined with delayed transport, and the fix is education of the procedural areas plus prompt pickup. Excess detergent does not create dried soil, and it would be observed after washing rather than on arrival. High humidity would tend to keep soil moist, and covering carts protects rather than harms the instruments.
法律依据: AAMI ST79免费刷完整 100 道题库 — 无需注册。
同考点相关题目
- Soiled instruments must be transported from the operating room to the decontamination area on a different floor. What is required?
- A decontamination area is equipped with three sinks. What is the intended use of each?
- A lumened suction instrument arrives in decontamination. Which cleaning practice is correct?
- A technician keeps a covered coffee cup on a shelf above the decontamination sink. How should this be addressed?
- Instruments are coming out of the washer with a chalky white film. Water testing shows high hardness. What is the appropriate corrective action?
- An enzymatic cleaner label lists protease, lipase, and amylase. What does this combination indicate?