Sterilization第 30 / 100 题
A wrapped instrument set is being sterilized in a dynamic-air-removal steam sterilizer at 270 degrees F (132 degrees C). What is the typical minimum exposure time?
a.1 minute
b.4 minutes
c.15 minutes
d.30 minutes
解析
Four minutes at 270 degrees F is the commonly validated exposure time for wrapped goods in a dynamic-air-removal cycle because air is removed before the exposure phase begins. Fifteen minutes at 270 degrees F and thirty minutes at 250 degrees F are gravity cycle parameters, which take longer because air removal is passive. One minute is not a validated exposure time for any wrapped steam cycle, and the device manufacturer instructions always govern the final selection.
法律依据: AAMI ST79免费刷完整 100 道题库 — 无需注册。
同考点相关题目
- What is the fundamental difference between a gravity displacement cycle and a dynamic-air-removal (prevacuum) cycle?
- What is a standard gravity displacement steam cycle for a wrapped instrument set at 250 degrees F (121 degrees C)?
- A technician opens the department at the start of the day shift and must schedule the Bowie-Dick test for a dynamic-air-removal sterilizer. How often is the test required?
- A Bowie-Dick test sheet comes out of the sterilizer with a pale, uneven center and darker edges. What does this indicate and what should be done?
- After a tray fails its internal chemical indicator, the supervisor reviews the fundamentals with the team. Which four parameters must all be satisfied for steam sterilization to be effective?
- Why is saturated steam, rather than superheated or wet steam, required for steam sterilization?