Prep, Packaging & Storage第 81 / 100 题
A peel pouch is being prepared for a single ring-handled instrument. Which practice is correct?
a.Select a pouch size that lets the item fit without stressing the seals, place the item so the handle enters the pouch first, and do not double pouch unless the pouch manufacturer validates it
b.Fill the pouch completely so the instrument cannot shift during handling
c.Fold the pouch over to fit a long instrument and seal it with tape
d.Write the contents on the paper side with a ballpoint pen
解析
A properly sized pouch keeps the seals intact, permits steam entry through the porous side, and allows the package to be opened aseptically with the handle presented first. Overfilling stresses seals, and folding or taping a pouch prevents sterilant contact and creates an unvalidated package. Labeling is done on the plastic side, outside the seal, with a non-bleeding marker, because writing on the paper side can damage the sterile barrier.
法律依据: AAMI ST79免费刷完整 100 道题库 — 无需注册。
同考点相关题目
- How should the cutting performance of surgical scissors be verified during inspection?
- A technician notices that a hinged instrument is stiff during inspection and asks which lubricant to apply before packaging. What is correct?
- An instrument set is being assembled. Which practice supports effective sterilization and safe handling?
- What must be verified about a rigid sterilization container each time it is used?
- A technician is applying labels to packages as they are assembled for sterilization. What information belongs on the label of a sterilized package?
- A facility uses event-related sterility. What does this mean?