Sterilization第 38 / 100 题
A tray is removed from the sterilizer and the outer wrapper feels damp to the touch. What is the correct action?
a.Set the tray aside until the moisture evaporates, then send it to storage
b.Consider the package contaminated, do not use or store it, and reprocess the contents after investigating the cause
c.Wipe the wrapper dry with a clean lint-free towel and release the tray
d.Place the tray back in the sterilizer for an additional drying cycle and then release it
解析
Moisture provides a wicking pathway for microorganisms from the outer surface to the contents, so a wet pack is considered contaminated regardless of whether the cycle parameters were met. The contents must be rewrapped and reprocessed, and the cause must be investigated. Air drying, wiping, or re-running only the drying phase does not restore sterility to a package that has already wicked.
法律依据: AAMI ST79免费刷完整 100 道题库 — 无需注册。
同考点相关题目
- Why is saturated steam, rather than superheated or wet steam, required for steam sterilization?
- Peel pouches are being loaded onto a sterilizer shelf. What is the correct placement?
- A mixed load contains wrapped metal basin sets and textile packs. How should the load be arranged?
- Wet packs are appearing intermittently across several loads. Which of the following is the least likely cause?
- After a steam cycle finishes and the load is removed, how should the load be handled?
- Immediate-use steam sterilization (IUSS) is requested for a single dropped instrument during a case. Which statement is correct?