Sterilization第 40 / 100 题
After a steam cycle finishes and the load is removed, how should the load be handled?
a.Transferred immediately to sterile storage to protect it from room air
b.Placed on a solid countertop and covered with a clean towel while it cools
c.Unloaded and immediately handled by staff wearing clean gloves
d.Left undisturbed on the cart in a low-traffic, draft-free area until fully cool, typically 30 minutes or longer, before it is touched or moved
解析
Hot packages are vulnerable because handling them can drive condensation and contamination through the packaging, so the load is left to cool on the cart away from vents, drafts, and traffic. Cooling times of 30 minutes to an hour or more are common depending on load density. Placing hot items on a cold solid surface or covering them traps condensate, and gloves do not protect a package that is still hot and damp.
法律依据: AAMI ST79免费刷完整 100 道题库 — 无需注册。
同考点相关题目
- A mixed load contains wrapped metal basin sets and textile packs. How should the load be arranged?
- A tray is removed from the sterilizer and the outer wrapper feels damp to the touch. What is the correct action?
- Wet packs are appearing intermittently across several loads. Which of the following is the least likely cause?
- Immediate-use steam sterilization (IUSS) is requested for a single dropped instrument during a case. Which statement is correct?
- Which practice regarding IUSS and implants is correct?
- A typical IUSS gravity cycle for nonporous metal instruments only is 270 degrees F (132 degrees C) for 3 minutes. What changes when the item is porous, cannulated, or has a lumen?