Sterilization第 42 / 100 题
Which practice regarding IUSS and implants is correct?
a.Implants may be routinely processed by IUSS as long as the case is urgent
b.IUSS of implants should be avoided; if unavoidable in a documented emergency, a rapid-readout biological indicator with a Class 5 integrating indicator is run and the event is documented
c.Implants may be processed by IUSS if a Class 1 process indicator is placed on the tray
d.Implants processed by IUSS require no biological monitoring because the cycle is unwrapped
解析
Because a failed implant sterilization carries severe patient consequences, implants should be sterilized in a full terminal cycle and quarantined until the biological indicator result is known. In a documented emergency, IUSS may be used with a process challenge device containing a rapid-readout biological indicator and a Class 5 integrating indicator, and the exception must be recorded. Convenience or urgency alone does not justify the practice, and an external process indicator only shows exposure.
法律依据: AAMI ST79免费刷完整 100 道题库 — 无需注册。
同考点相关题目
- Wet packs are appearing intermittently across several loads. Which of the following is the least likely cause?
- After a steam cycle finishes and the load is removed, how should the load be handled?
- Immediate-use steam sterilization (IUSS) is requested for a single dropped instrument during a case. Which statement is correct?
- A typical IUSS gravity cycle for nonporous metal instruments only is 270 degrees F (132 degrees C) for 3 minutes. What changes when the item is porous, cannulated, or has a lumen?