A technician is loading a steam sterilizer and wants to fit as many trays as possible before the shift ends. How full should the chamber be loaded?
a.Completely full, with items packed tightly so nothing shifts during the cycle
b.At least three quarters full, because partial loads waste energy
c.Loosely, with space between packages and between packages and the chamber walls so steam can circulate freely
d.Only with identical items, never mixed loads, regardless of spacing
解析
Free circulation of steam and unobstructed air removal require space around every package, so overloading is a leading cause of sterilization failure and wet packs. Packing tightly or filling to a minimum percentage for efficiency compromises the process. Mixed loads are acceptable when items are arranged correctly, for example metal below textiles, and when all items share the same validated cycle.
法律依据: AAMI ST79免费刷完整 100 道题库 — 无需注册。
同考点相关题目
- A technician notices the sterilizer printout shows the chamber reached only 268 degrees F during a cycle set for 270 degrees F, although the chemical indicator changed color. What should be done?
- Why must instruments with hinges, such as hemostats, be placed in the sterilizer in the open, unlocked position?
- A student asks a technician what it means when a process is described as achieving a sterility assurance level of 10 to the minus 6. What is the correct explanation?
- A liquid chemical sterilant such as peracetic acid is used to process a device. What is the key limitation of this method?
- An instrument manufacturer's instructions for use specify a prevacuum cycle at 270 degrees F for 4 minutes with an 8-minute dry time, but department policy uses a 30-minute dry time for all trays. Which should be followed?
- During a steam cycle the sterilizer aborts and displays an air leak alarm. What is the correct response?