Sterilization第 53 / 100 题
Why must instruments with hinges, such as hemostats, be placed in the sterilizer in the open, unlocked position?
a.Steam must contact every surface, and closed box locks and ratchets shield the mating surfaces from the sterilant
b.Closed instruments trap air that lowers the chamber pressure
c.Open instruments dry faster and reduce total cycle time
d.Closed instruments will be damaged by the pressure changes in the chamber
解析
Sterilization requires direct sterilant contact with all surfaces, and a closed ratchet creates a shielded metal-to-metal interface that steam cannot reach. Stringers, racks, and instrument pins are used to hold hinged instruments open throughout the cycle. Chamber pressure is controlled by the sterilizer, drying benefit is secondary, and correctly manufactured instruments are not damaged by normal cycle pressure changes.
法律依据: AAMI ST79免费刷完整 100 道题库 — 无需注册。
同考点相关题目
- A technician receives a jar of petroleum-based lubricant and a container of powder that must be sterilized. Which method applies, and what are representative parameters?
- Rigid sterilization containers are being loaded into a steam sterilizer along with wrapped trays. What is correct practice?
- A technician notices the sterilizer printout shows the chamber reached only 268 degrees F during a cycle set for 270 degrees F, although the chemical indicator changed color. What should be done?
- A student asks a technician what it means when a process is described as achieving a sterility assurance level of 10 to the minus 6. What is the correct explanation?
- A technician is loading a steam sterilizer and wants to fit as many trays as possible before the shift ends. How full should the chamber be loaded?
- A liquid chemical sterilant such as peracetic acid is used to process a device. What is the key limitation of this method?