A technician receives a jar of petroleum-based lubricant and a container of powder that must be sterilized. Which method applies, and what are representative parameters?
a.Flexible endoscopes, at 250 degrees F for 30 minutes
b.Oils, petroleum-based lubricants, and powders, at approximately 320 degrees F for 2 hours or 340 degrees F for 1 hour
c.Rigid containers with filters, at 270 degrees F for 4 minutes
d.Textile packs, at 275 degrees F for 3 minutes
解析
Dry heat is used for anhydrous materials that steam cannot penetrate, such as oils, ointments, and powders, and it needs long exposure at high temperature because heat transfer without moisture is inefficient. Flexible endoscopes are heat sensitive and are high-level disinfected or low-temperature sterilized. The 270 and 275 degrees F short cycles listed are steam parameters, not dry heat.
法律依据: AAMI ST79免费刷完整 100 道题库 — 无需注册。
同考点相关题目
- What are the byproducts of a hydrogen peroxide gas plasma sterilization cycle, and what does this mean operationally?
- An ozone sterilization system is being introduced. Which statement about ozone sterilization is accurate?
- A department is deciding which sterilization method to use for a heat-stable stainless steel instrument set. Which method should be selected and why?
- Rigid sterilization containers are being loaded into a steam sterilizer along with wrapped trays. What is correct practice?
- A technician notices the sterilizer printout shows the chamber reached only 268 degrees F during a cycle set for 270 degrees F, although the chemical indicator changed color. What should be done?
- Why must instruments with hinges, such as hemostats, be placed in the sterilizer in the open, unlocked position?