Monitoring & QA第 65 / 100 题

Where should the process challenge device be placed in a steam sterilizer chamber for routine biological monitoring?

a.On the top shelf at the rear of the chamber
b.In the geometric center of the load, surrounded by the densest packs
c.In the area identified as the most difficult for the sterilant to reach, typically on the bottom shelf near the front over the drain
d.Outside the chamber, taped to the sterilizer door

解析

Air is heaviest and settles low near the drain, so the front bottom of the chamber is the recognized cold spot for a steam sterilizer and is where the PCD is placed. Testing an easy location would give a falsely reassuring result. The exact placement follows the sterilizer manufacturer instructions, and an indicator outside the chamber tests nothing.

法律依据: AAMI ST79

免费刷完整 100 道题库 — 无需注册。

同考点相关题目

最近核对: · 审核流程

PrepPass 编辑团队 · 依据官方资料核对 HSPA Certified Registered Central Service Technician (CRCST) Exam · 我们如何核对