Monitoring & QA第 67 / 100 题

A biological indicator from this morning's steam load is positive. What is the first action?

a.Take the sterilizer out of service and immediately recall and quarantine the items from that load and from every load back to the last negative biological indicator
b.Repeat the test at the end of the shift and continue processing loads in the meantime
c.Discard the indicator as a false positive, since spore vials are unreliable
d.Notify the operating room that items may have been unsterile but allow already distributed items to be used

解析

A positive biological indicator means the process cannot be assumed to have been lethal, so the sterilizer is removed from service and every load processed since the last negative result is retrieved, quarantined, and reprocessed. Continuing to run loads or waiting to retest extends the exposure to patients. Positives are investigated, not assumed to be false, and previously distributed items must be located, not released for use.

法律依据: AAMI ST79