Monitoring & QA第 73 / 100 题

What is the correct handling of the control biological indicator that accompanies a test lot?

a.It is placed in the sterilizer alongside the test indicator
b.It is discarded because it is not part of the record
c.It is not sterilized, is incubated with the processed indicator from the same lot, and must show growth for the test to be valid
d.It is stored in a refrigerator until the end of the month

解析

The unprocessed control proves the spore population is viable and that the incubator and reading method are working, so it must turn positive while the processed indicator turns negative. If the control fails to grow, the test result is invalid regardless of what the processed indicator shows. Sterilizing the control defeats its purpose, and it must come from the same lot and be incubated at the same time.

法律依据: AAMI ST79

免费刷完整 100 道题库 — 无需注册。

同考点相关题目

最近核对: · 审核流程

PrepPass 编辑团队 · 依据官方资料核对 HSPA Certified Registered Central Service Technician (CRCST) Exam · 我们如何核对