DecontaminationQuestion 23 of 100
Instruments are coming out of the washer with a chalky white film. Water testing shows high hardness. What is the appropriate corrective action?
a.Increase the alkaline detergent concentration to dissolve the film
b.Extend the wash phase by ten minutes on every cycle
c.Address water quality with treatment such as softening or deionization, and use a descaling product as directed for existing deposits
d.Lower the final rinse temperature so minerals stay in solution
Explanation
A white chalky film is mineral scale from hard water depositing during rinsing and drying, and the durable fix is treating the water supply, with an acidic descaler used per manufacturer instructions to remove existing buildup. Adding alkaline detergent does not dissolve mineral scale and risks damaging aluminum. Longer wash times and cooler rinses do not change the mineral content of the water.
Law Reference: AAMI ST79Practice all 100 questions free — no signup required.
Related questions on this topic
- A lumened suction instrument arrives in decontamination. Which cleaning practice is correct?
- Instrument sets are consistently arriving in decontamination with hard, dried blood in box locks and serrations. What is the most likely root cause?
- A technician keeps a covered coffee cup on a shelf above the decontamination sink. How should this be addressed?
- An enzymatic cleaner label lists protease, lipase, and amylase. What does this combination indicate?
- A technician finishes cleaning and is about to walk into the clean assembly area. What must occur?
- A case cart cannot be picked up for several hours because of staffing. What is the best interim measure for the soiled instruments?
Last reviewed: · editorial process
PrepPass Editorial Team · Verified against HSPA Certified Registered Central Service Technician (CRCST) Exam · How we review