A typical IUSS gravity cycle for nonporous metal instruments only is 270 degrees F (132 degrees C) for 3 minutes. What changes when the item is porous, cannulated, or has a lumen?
a.The temperature is lowered to 250 degrees F and the time stays at 3 minutes
b.The cycle is unchanged because IUSS parameters are fixed
c.A longer exposure, commonly 10 minutes in a gravity cycle, is required per the device manufacturer instructions
d.The item cannot be sterilized by steam under any circumstance
Explanation
Porous items, lumens, and complex devices are harder for steam to penetrate, so validated IUSS parameters extend the exposure time, commonly to 10 minutes in a gravity cycle. The device manufacturer written instructions for use always define the required cycle. Lowering the temperature reduces lethality, IUSS parameters are not fixed across device types, and most such devices are perfectly steam compatible.
Law Reference: AAMI ST79Practice all 100 questions free — no signup required.
Related questions on this topic
- After a steam cycle finishes and the load is removed, how should the load be handled?
- Immediate-use steam sterilization (IUSS) is requested for a single dropped instrument during a case. Which statement is correct?
- Which practice regarding IUSS and implants is correct?
- A surgeon sends a heat-sensitive device with a long narrow lumen and asks whether it can be processed with ethylene oxide. Which statement describes ethylene oxide sterilization?
- Why must every ethylene oxide load undergo an aeration phase before items are released?
- A tray wrapped in a cellulose-based (paper) wrapper is about to be loaded into a hydrogen peroxide gas plasma sterilizer. What is the problem?
Last reviewed: · editorial process
PrepPass Editorial Team · Verified against HSPA Certified Registered Central Service Technician (CRCST) Exam · How we review