A technician is scrubbing a hinged instrument with a brush at the decontamination sink and is spraying water droplets into the air. What is the correct technique?
a.Brush more slowly under a running faucet so droplets fall back into the sink
b.Keep the instrument and the brush fully submerged beneath the water surface while brushing
c.Brush above the water line so soil is visible and can be rinsed away immediately
d.Substitute a high-pressure water gun for the brush at all times
解析
Brushing under the surface of the water contains splash and prevents the release of contaminated aerosols that expose the worker and the room air. Working above the water line, even slowly under a faucet, still generates droplets. Pressurized water guns have their place for lumens but must be used with additional splash protection and do not replace submerged brushing.
法律依据: AAMI ST79免费刷完整 100 道题库 — 无需注册。
同考点相关题目
- A technician fills the wash sink with very hot water to speed up cleaning. Why must manual cleaning water stay below approximately 140 degrees F (60 degrees C)?
- Most instrument manufacturers specify a neutral pH detergent for routine cleaning. What pH range is considered neutral?
- A washer-disinfector is being validated and the question arises about which water quality is required for the final rinse. What is correct?
- What is the purpose of the elevated-temperature rinse phase, often around 180 degrees F or higher, in a washer-disinfector cycle?
- A department is writing its quality plan and must set a frequency for cleaning verification testing of its mechanical washers. What should the plan specify?
- A new sterile processing department is being designed. Which statement describes correct workflow?