Decontamination第 15 / 100 题

What is the purpose of the elevated-temperature rinse phase, often around 180 degrees F or higher, in a washer-disinfector cycle?

a.It sterilizes the instruments so they can bypass terminal sterilization
b.It activates the enzymatic detergent so proteins continue to break down
c.It provides thermal disinfection that reduces microbial load and improves drying
d.It cools the load gradually to prevent thermal shock to delicate instruments

解析

The hot rinse delivers thermal disinfection, lowering the bioburden so items can be safely handled in the clean workroom, and the residual heat helps the load dry. Thermal disinfection is not sterilization, so terminal sterilization is still required for critical devices. Enzymes are destroyed at that temperature, and the phase heats rather than cools the load.

法律依据: AAMI ST79

免费刷完整 100 道题库 — 无需注册。

同考点相关题目

最近核对: · 审核流程

PrepPass 编辑团队 · 依据官方资料核对 HSPA Certified Registered Central Service Technician (CRCST) Exam · 我们如何核对