Why should the lid of an ultrasonic cleaner remain closed while the unit is running?
a.To maintain the water temperature required for thermal disinfection
b.To prevent the detergent from evaporating and changing concentration
c.To contain the contaminated aerosols that the cavitation process releases into the air
d.To keep instruments from floating out of the basket
Explanation
Cavitation drives contaminated droplets off the solution surface, so keeping the lid closed protects staff and the room air from aerosolized bioburden. Ultrasonic cleaners do not perform thermal disinfection. Evaporation and floating instruments are minor operational issues, not the safety-driven reason for the closed lid.
Law Reference: AAMI ST79Practice all 100 questions free — no signup required.
Related questions on this topic
- A technician keeps a covered coffee cup on a shelf above the decontamination sink. How should this be addressed?
- Instruments are coming out of the washer with a chalky white film. Water testing shows high hardness. What is the appropriate corrective action?
- An enzymatic cleaner label lists protease, lipase, and amylase. What does this combination indicate?
- A technician finishes cleaning and is about to walk into the clean assembly area. What must occur?
- A case cart cannot be picked up for several hours because of staffing. What is the best interim measure for the soiled instruments?
- During a competency review, a technician is asked to explain why cleaning is considered the single most important step in instrument reprocessing. Which answer is correct?
Last reviewed: · editorial process
PrepPass Editorial Team · Verified against HSPA Certified Registered Central Service Technician (CRCST) Exam · How we review