A recall requires the department to trace every package from a suspect cycle. What information must the load record contain?
a.Only the date and the technician initials
b.Only the sterilizer number and the cycle type
c.Only a list of the departments the items were sent to
d.Sterilizer identification, cycle or load number, date and time, contents of the load, cycle parameters, operator identification, and the results of physical, chemical, and biological monitoring
Explanation
Complete load documentation permits a targeted recall, links a specific package to a specific cycle, and demonstrates that monitoring was performed and reviewed. Each package also carries a lot control label with the sterilizer, cycle, and date so it can be matched back to the record. Partial records leave gaps that force a much broader and more disruptive recall.
Law Reference: AAMI ST79Practice all 100 questions free — no signup required.
Related questions on this topic
- A biological indicator from this morning's steam load is positive. What is the first action?
- After a positive biological indicator, the sterilizer is inspected, a probable cause is corrected, and testing resumes. What is generally required before the sterilizer returns to routine service?
- An implant is processed in a terminal steam cycle with a rapid-readout biological indicator. When may the implant be released?
- A nurse reports that a tray opened in the operating room contained an internal chemical indicator that did not change. What is the correct action?
- How should the daily Bowie-Dick test cycle be run?
- What is the correct handling of the control biological indicator that accompanies a test lot?
Last reviewed: · editorial process
PrepPass Editorial Team · Verified against HSPA Certified Registered Central Service Technician (CRCST) Exam · How we review