Monitoring & QAQuestion 72 of 100

How should the daily Bowie-Dick test cycle be run?

a.In a fully loaded chamber, to reflect real conditions
b.In an otherwise empty chamber, in the first cycle of the day after a warm-up cycle, at 270 to 275 degrees F for the time specified by the test manufacturer, usually not exceeding 4 minutes
c.In any cycle of the day, as long as the result is recorded
d.Together with a biological indicator, so both tests share the same cycle

Explanation

The Bowie-Dick test is run in an empty chamber so the only challenge to air removal is the test pack itself, and a warm-up cycle first brings the chamber and jacket to operating condition. Exposure time follows the test manufacturer instructions and is typically limited to about 3.5 to 4 minutes, because overexposure can produce a false pass. Running it in a production load or combining it with biological monitoring invalidates the intent of the test.

Law Reference: AAMI ST79

Practice all 100 questions free — no signup required.

Related questions on this topic

Last reviewed: · editorial process

PrepPass Editorial Team · Verified against HSPA Certified Registered Central Service Technician (CRCST) Exam · How we review