EKG AcquisitionCâu 40 / 100
A technician is placing V5 on a 12-lead EKG. Which description is correct?
a.Fifth intercostal space, left midaxillary line
b.Sixth intercostal space, left anterior axillary line
c.Fourth intercostal space, left anterior axillary line
d.Left anterior axillary line, on the same horizontal level as V4
Giải thích
V5 is placed at the left anterior axillary line on the same horizontal plane as V4 rather than by counting rib spaces, because the chest wall curves and the horizontal alignment keeps V4, V5, and V6 comparable. The midaxillary line at that level is V6. Counting to the sixth space or the fourth space would place the electrode too low or too high.
Luyện miễn phí toàn bộ 100 câu hỏi — không cần đăng ký.
Câu hỏi liên quan cùng chủ đề
- A technician has correctly placed V1. Where does the V2 electrode go?
- Which chest electrode is placed in the fifth intercostal space at the left midclavicular line?
- After placing V1, V2, and V4, where should the technician place V3?
- Where is the V6 electrode placed?
- Where should the limb electrodes be placed for a standard 12-lead EKG?
- A 12-lead tracing shows an inverted P wave, inverted QRS, and inverted T wave in lead I, while the chest leads look normal. What is the most likely cause?