EKG AcquisitionCâu 53 / 100
The QRS complexes on a 12-lead tracing are so tall that they run off the paper and overlap the lead above. What is the appropriate action?
a.Increase the paper speed to 50 mm/sec
b.Change the standardization to half sensitivity (5 mm/mV) and document the change on the tracing
c.Move the chest electrodes one intercostal space lower
d.Reduce the gain until the complexes look like the patient's last EKG
Giải thích
When complexes are too tall to fit, the machine is switched to half standardization, so 1 millivolt produces a 5 millimeter deflection, and the change must be marked on the tracing so the physician can interpret amplitudes correctly. Increasing paper speed widens complexes but does not lower their height. Moving electrodes falsifies the anatomic view, and adjusting the gain to match a prior tracing is not an accepted practice.
Luyện miễn phí toàn bộ 100 câu hỏi — không cần đăng ký.
Câu hỏi liên quan cùng chủ đề
- A patient has a hairy chest and oily skin. What is the correct skin preparation before applying chest electrodes?
- A technician is setting up 5-lead telemetry monitoring. Which placement follows the standard color convention?
- How many electrodes are applied for standard 3-lead cardiac monitoring, and what do they produce?
- A patient's heart rate is so fast that the complexes are crowded together and hard to measure. Which machine adjustment can help, and what must the technician remember?
- What is the standard patient position for obtaining a resting 12-lead EKG?
- A patient with severe shortness of breath cannot tolerate lying flat for a 12-lead EKG. What should the technician do?