SterilizationCâu 46 / 100

A tray wrapped in a cellulose-based (paper) wrapper is about to be loaded into a hydrogen peroxide gas plasma sterilizer. What is the problem?

a.Cellulose reacts with hydrogen peroxide to form a toxic residue on instruments
b.Cellulose absorbs hydrogen peroxide and will cause the cycle to abort or fail, so only nonwoven polypropylene wrap and compatible trays may be used
c.Cellulose wrap melts at the operating temperature of the sterilizer
d.Cellulose wrap is acceptable if a second wrapper is applied over it

Giải thích

Hydrogen peroxide is readily absorbed by cellulose materials such as paper, linen, and cotton, which strips sterilant from the chamber and causes cycle cancellation, so only polypropylene wrap, compatible pouches, and approved containers are used. The issue is absorption, not toxic residue formation or melting, since the process runs at low temperature. Adding a second cellulose layer makes the absorption problem worse.

Trích dẫn luật: AAMI ST79

Luyện miễn phí toàn bộ 100 câu hỏi — không cần đăng ký.

Câu hỏi liên quan cùng chủ đề

Cập nhật gần nhất: · quy trình kiểm tra

Đội Ngũ Biên Tập PrepPass · Đối chiếu với HSPA Certified Registered Central Service Technician (CRCST) Exam · Quy trình kiểm tra