Monitoring & QACâu 63 / 100
A technician finds a box of chemical indicators labeled for use only in a cycle of 270 degrees F for 4 minutes. What type of indicator is this?
a.An emulating indicator designed and verified for one specific cycle, reacting to all critical variables at the stated values of that cycle
b.An indicator used only in ethylene oxide sterilizers
c.An indicator that replaces the daily biological indicator
d.An external indicator used to identify the sterilizer number
Giải thích
Class 6 emulating indicators are cycle specific, meaning an indicator labeled for 270 degrees F and 4 minutes may be used only for that cycle, and they respond to all critical variables at the stated values. They exist for multiple sterilization methods, not just ethylene oxide. No chemical indicator of any class replaces biological monitoring, and identifying the sterilizer is a labeling function.
Trích dẫn luật: AAMI ST79Luyện miễn phí toàn bộ 100 câu hỏi — không cần đăng ký.
Câu hỏi liên quan cùng chủ đề
- A department is setting its biological monitoring policy for a steam sterilizer used for routine loads. How frequently should a biological indicator be run?
- A Class 1 chemical indicator is applied to the outside of every wrapped tray. What does it tell the user?
- Which class of chemical indicator reacts to all critical variables of the sterilization cycle and has performance correlated to a biological indicator?
- A supervisor asks a technician to assemble a process challenge device for the daily biological indicator test. What is a process challenge device?
- Where should the process challenge device be placed in a steam sterilizer chamber for routine biological monitoring?
- Which of the following are physical monitors of a sterilization cycle?
Cập nhật gần nhất: · quy trình kiểm tra
Đội Ngũ Biên Tập PrepPass · Đối chiếu với HSPA Certified Registered Central Service Technician (CRCST) Exam · Quy trình kiểm tra