Prep, Packaging & StorageCâu 82 / 100
What must be verified about a rigid sterilization container each time it is used?
a.That the container has been polished and is free of fingerprints
b.That filters are correct, unused, and properly seated; that valves, gaskets, latches, and the lid seal are intact; and that a tamper-evident lock is applied after assembly
c.That the container is completely airtight so no organisms can enter
d.That the container weighs less than 5 pounds when empty
Giải thích
Rigid containers depend on intact filters, gaskets, and valves to allow sterilant in and keep contaminants out, so each of these is inspected before every use and single-use filters are replaced. A tamper-evident lock shows at the point of use that the container has not been opened since processing. Containers are deliberately not airtight, since steam must enter, and empty weight is not a use criterion.
Trích dẫn luật: AAMI ST79Luyện miễn phí toàn bộ 100 câu hỏi — không cần đăng ký.
Câu hỏi liên quan cùng chủ đề
- A technician notices that a hinged instrument is stiff during inspection and asks which lubricant to apply before packaging. What is correct?
- An instrument set is being assembled. Which practice supports effective sterilization and safe handling?
- A peel pouch is being prepared for a single ring-handled instrument. Which practice is correct?
- A technician is applying labels to packages as they are assembled for sterilization. What information belongs on the label of a sterilized package?
- A facility uses event-related sterility. What does this mean?
- What temperature and humidity conditions are recommended for a sterile storage area?
Cập nhật gần nhất: · quy trình kiểm tra
Đội Ngũ Biên Tập PrepPass · Đối chiếu với HSPA Certified Registered Central Service Technician (CRCST) Exam · Quy trình kiểm tra