DecontaminationCâu 15 / 100
What is the purpose of the elevated-temperature rinse phase, often around 180 degrees F or higher, in a washer-disinfector cycle?
a.It sterilizes the instruments so they can bypass terminal sterilization
b.It activates the enzymatic detergent so proteins continue to break down
c.It provides thermal disinfection that reduces microbial load and improves drying
d.It cools the load gradually to prevent thermal shock to delicate instruments
Giải thích
The hot rinse delivers thermal disinfection, lowering the bioburden so items can be safely handled in the clean workroom, and the residual heat helps the load dry. Thermal disinfection is not sterilization, so terminal sterilization is still required for critical devices. Enzymes are destroyed at that temperature, and the phase heats rather than cools the load.
Trích dẫn luật: AAMI ST79Luyện miễn phí toàn bộ 100 câu hỏi — không cần đăng ký.
Câu hỏi liên quan cùng chủ đề
- Most instrument manufacturers specify a neutral pH detergent for routine cleaning. What pH range is considered neutral?
- A washer-disinfector is being validated and the question arises about which water quality is required for the final rinse. What is correct?
- A technician is scrubbing a hinged instrument with a brush at the decontamination sink and is spraying water droplets into the air. What is the correct technique?
- A department is writing its quality plan and must set a frequency for cleaning verification testing of its mechanical washers. What should the plan specify?
- A new sterile processing department is being designed. Which statement describes correct workflow?
- Soiled instruments must be transported from the operating room to the decontamination area on a different floor. What is required?
Cập nhật gần nhất: · quy trình kiểm tra
Đội Ngũ Biên Tập PrepPass · Đối chiếu với HSPA Certified Registered Central Service Technician (CRCST) Exam · Quy trình kiểm tra