A department is writing its quality plan and must set a frequency for cleaning verification testing of its mechanical washers. What should the plan specify?
a.Only when a visible soil complaint is reported by the operating room
b.Once per quarter, at the same time as preventive maintenance
c.Once per month, at the start of the month
d.At least weekly, and preferably each day the washer is used
Giải thích
Routine cleaning verification with a commercial soil test device confirms the washer is still removing soil, and the recommendation is at least weekly with daily testing preferred. Waiting for a complaint is reactive and allows contaminated devices to reach patients. Quarterly and monthly intervals are too infrequent to catch a spray arm blockage, pump failure, or chemical delivery problem in time.
Trích dẫn luật: AAMI ST79Luyện miễn phí toàn bộ 100 câu hỏi — không cần đăng ký.
Câu hỏi liên quan cùng chủ đề
- A washer-disinfector is being validated and the question arises about which water quality is required for the final rinse. What is correct?
- A technician is scrubbing a hinged instrument with a brush at the decontamination sink and is spraying water droplets into the air. What is the correct technique?
- What is the purpose of the elevated-temperature rinse phase, often around 180 degrees F or higher, in a washer-disinfector cycle?
- A new sterile processing department is being designed. Which statement describes correct workflow?
- Soiled instruments must be transported from the operating room to the decontamination area on a different floor. What is required?
- A decontamination area is equipped with three sinks. What is the intended use of each?
Cập nhật gần nhất: · quy trình kiểm tra
Đội Ngũ Biên Tập PrepPass · Đối chiếu với HSPA Certified Registered Central Service Technician (CRCST) Exam · Quy trình kiểm tra