EKG AcquisitionPregunta 61 de 100
A technician opens a package of disposable electrodes and finds the conductive gel is dry and crusted. What should be done?
a.Moisten the electrodes with tap water and use them
b.Discard the package and open fresh electrodes within their expiration date
c.Add alcohol to the gel to soften it
d.Use them anyway and increase the machine gain to compensate
Explicación
Dried electrodes cannot conduct properly and cause wandering baseline and false readings, so they are discarded and replaced with fresh, unexpired electrodes stored in a sealed package. Water and alcohol are not substitutes for conductive gel and alcohol actually increases resistance once it dries. Increasing gain amplifies the artifact along with the signal and does not fix the contact problem.
Practica las 100 preguntas gratis — sin registro.
Preguntas relacionadas de este tema
- A physician orders right-sided chest leads on a patient with a suspected right ventricular infarction. Which electrode position corresponds to V4R?
- A patient has a right below-the-knee amputation. How should the technician place the limb electrodes?
- When performing a 12-lead EKG on a female patient with large breast tissue, where should the V4 and V5 electrodes be placed?
- On a 12-lead tracing, leads I, II, and aVR are clean, but lead III alone shows an erratic, noisy trace. What is the most likely cause?
- A tracing shows steady 60-cycle interference. Which troubleshooting sequence is most appropriate?
- Before beginning a shift, a technician inspects the EKG machine. Which finding requires the machine be removed from service?
Última revisión: · proceso editorial
Equipo Editorial de PrepPass · Verificado con NHA Certified EKG Technician (CET) Exam · Cómo revisamos