SterilizationPregunta 35 de 100
Why is saturated steam, rather than superheated or wet steam, required for steam sterilization?
a.Superheated steam is too cold to kill spores
b.Wet steam carries more heat energy and therefore sterilizes faster
c.Saturated steam gives up its latent heat on contact with cooler surfaces, and both superheat and excess moisture disrupt that heat transfer
d.Only saturated steam can pass through a sterilization wrapper
Explicación
Sterilization depends on condensation, which releases latent heat directly onto the item surface, and saturated steam near 97 percent dryness delivers that transfer efficiently. Superheated steam behaves like hot dry air and requires far longer exposure times, while wet steam carries excess water that causes wet packs and reduced penetration. Steam of any quality can physically pass through a wrapper, so that is not the distinguishing factor.
Referencia Legal: AAMI ST79Practica las 100 preguntas gratis — sin registro.
Preguntas relacionadas de este tema
- A technician opens the department at the start of the day shift and must schedule the Bowie-Dick test for a dynamic-air-removal sterilizer. How often is the test required?
- A Bowie-Dick test sheet comes out of the sterilizer with a pale, uneven center and darker edges. What does this indicate and what should be done?
- After a tray fails its internal chemical indicator, the supervisor reviews the fundamentals with the team. Which four parameters must all be satisfied for steam sterilization to be effective?
- Peel pouches are being loaded onto a sterilizer shelf. What is the correct placement?
- A mixed load contains wrapped metal basin sets and textile packs. How should the load be arranged?
- A tray is removed from the sterilizer and the outer wrapper feels damp to the touch. What is the correct action?
Última revisión: · proceso editorial
Equipo Editorial de PrepPass · Verificado con HSPA Certified Registered Central Service Technician (CRCST) Exam · Cómo revisamos