Prep, Packaging & StoragePregunta 78 de 100
How should the cutting performance of surgical scissors be verified during inspection?
a.By running a fingernail along the blade edge
b.By cutting through the appropriate test material, such as several layers of gauze or a test material specified by the manufacturer, using the distal third of the blades
c.By holding the blades up to the light to check for gaps
d.By opening and closing the scissors to check the pivot screw only
Explicación
Scissors are tested by cutting a standardized material with the tips, because the distal portion of the blades is where dullness first appears and where clinical cutting failures occur. Feeling the edge is subjective and creates a sharps injury risk. Light inspection and pivot checks are useful additional steps but do not confirm cutting performance.
Referencia Legal: AAMI ST79Practica las 100 preguntas gratis — sin registro.
Preguntas relacionadas de este tema
- A technician is inspecting instruments at the assembly workstation. Which tool and technique should be used?
- A technician notices that a hinged instrument is stiff during inspection and asks which lubricant to apply before packaging. What is correct?
- An instrument set is being assembled. Which practice supports effective sterilization and safe handling?
- A peel pouch is being prepared for a single ring-handled instrument. Which practice is correct?
- What must be verified about a rigid sterilization container each time it is used?
- A technician is applying labels to packages as they are assembled for sterilization. What information belongs on the label of a sterilized package?
Última revisión: · proceso editorial
Equipo Editorial de PrepPass · Verificado con HSPA Certified Registered Central Service Technician (CRCST) Exam · Cómo revisamos