SterilizationCâu 38 / 100

A tray is removed from the sterilizer and the outer wrapper feels damp to the touch. What is the correct action?

a.Set the tray aside until the moisture evaporates, then send it to storage
b.Consider the package contaminated, do not use or store it, and reprocess the contents after investigating the cause
c.Wipe the wrapper dry with a clean lint-free towel and release the tray
d.Place the tray back in the sterilizer for an additional drying cycle and then release it

Giải thích

Moisture provides a wicking pathway for microorganisms from the outer surface to the contents, so a wet pack is considered contaminated regardless of whether the cycle parameters were met. The contents must be rewrapped and reprocessed, and the cause must be investigated. Air drying, wiping, or re-running only the drying phase does not restore sterility to a package that has already wicked.

Trích dẫn luật: AAMI ST79

Luyện miễn phí toàn bộ 100 câu hỏi — không cần đăng ký.