SterilizationCâu 36 / 100
Peel pouches are being loaded onto a sterilizer shelf. What is the correct placement?
a.Stacked flat, paper side up, no more than ten deep
b.Stacked flat, plastic side up, so condensate runs off the film
c.Laid flat and overlapping to maximize the number of pouches per shelf
d.Positioned on edge in a rack, paper surface facing the plastic surface of the next pouch
Giải thích
Placing pouches on edge with paper facing plastic lets steam enter through the porous paper and lets condensate drain away, which also improves drying. Stacking pouches flat traps air and moisture between them and blocks the porous side. Overlapping pouches likewise prevents steam contact and is a common cause of wet packs and failed indicators.
Trích dẫn luật: AAMI ST79Luyện miễn phí toàn bộ 100 câu hỏi — không cần đăng ký.
Câu hỏi liên quan cùng chủ đề
- A Bowie-Dick test sheet comes out of the sterilizer with a pale, uneven center and darker edges. What does this indicate and what should be done?
- After a tray fails its internal chemical indicator, the supervisor reviews the fundamentals with the team. Which four parameters must all be satisfied for steam sterilization to be effective?
- Why is saturated steam, rather than superheated or wet steam, required for steam sterilization?
- A mixed load contains wrapped metal basin sets and textile packs. How should the load be arranged?
- A tray is removed from the sterilizer and the outer wrapper feels damp to the touch. What is the correct action?
- Wet packs are appearing intermittently across several loads. Which of the following is the least likely cause?