After a positive biological indicator, the sterilizer is inspected, a probable cause is corrected, and testing resumes. What is generally required before the sterilizer returns to routine service?
a.A single passing chemical indicator in a full production load
b.Three consecutive cycles with negative biological indicators in an empty chamber, and for dynamic-air-removal sterilizers three passing Bowie-Dick tests as well
c.One passing biological indicator run in a fully loaded chamber
d.A signed statement from the operator that the load appeared normal
Giải thích
Requalification after a failure or a major repair uses three consecutive empty-chamber cycles with negative biological indicators, with Bowie-Dick testing added for dynamic-air-removal sterilizers, to demonstrate the equipment is reliably performing again. A chemical indicator does not demonstrate microbial lethality, and one biological indicator in a production load is not sufficient evidence of restored performance. Operator observation is not a qualification method.
Trích dẫn luật: AAMI ST79Luyện miễn phí toàn bộ 100 câu hỏi — không cần đăng ký.
Câu hỏi liên quan cùng chủ đề
- Where should the process challenge device be placed in a steam sterilizer chamber for routine biological monitoring?
- Which of the following are physical monitors of a sterilization cycle?
- A biological indicator from this morning's steam load is positive. What is the first action?
- An implant is processed in a terminal steam cycle with a rapid-readout biological indicator. When may the implant be released?
- A recall requires the department to trace every package from a suspect cycle. What information must the load record contain?
- A nurse reports that a tray opened in the operating room contained an internal chemical indicator that did not change. What is the correct action?
Cập nhật gần nhất: · quy trình kiểm tra
Đội Ngũ Biên Tập PrepPass · Đối chiếu với HSPA Certified Registered Central Service Technician (CRCST) Exam · Quy trình kiểm tra