Monitoring & QACâu 71 / 100
A nurse reports that a tray opened in the operating room contained an internal chemical indicator that did not change. What is the correct action?
a.Do not use the tray, return it for reprocessing, and investigate the cycle records and the rest of the load
b.Use the tray if the external tape changed color, since two indicators are redundant
c.Use the tray but document the indicator failure afterward
d.Replace the indicator and reseal the tray in the operating room
Giải thích
An internal indicator that failed to change means sterilant may not have reached the interior of the package, so the tray is treated as unsterile and reprocessed while the load records and remaining packages are reviewed. External tape only shows the package was exposed to a process, so it cannot override an internal failure. Resealing or using the tray and documenting later places the patient at risk.
Trích dẫn luật: AAMI ST79Luyện miễn phí toàn bộ 100 câu hỏi — không cần đăng ký.
Câu hỏi liên quan cùng chủ đề
- After a positive biological indicator, the sterilizer is inspected, a probable cause is corrected, and testing resumes. What is generally required before the sterilizer returns to routine service?
- An implant is processed in a terminal steam cycle with a rapid-readout biological indicator. When may the implant be released?
- A recall requires the department to trace every package from a suspect cycle. What information must the load record contain?
- How should the daily Bowie-Dick test cycle be run?
- What is the correct handling of the control biological indicator that accompanies a test lot?
- Which statement best describes the regulatory and standards landscape for sterile processing?
Cập nhật gần nhất: · quy trình kiểm tra
Đội Ngũ Biên Tập PrepPass · Đối chiếu với HSPA Certified Registered Central Service Technician (CRCST) Exam · Quy trình kiểm tra