Monitoring & QACâu 69 / 100
An implant is processed in a terminal steam cycle with a rapid-readout biological indicator. When may the implant be released?
a.Immediately, if the Class 5 integrating indicator inside the tray has changed
b.At the end of the shift, regardless of the biological indicator result
c.Only after the biological indicator result is available and negative, except in a documented emergency
d.Only after 48 hours of incubation, with no exceptions
Giải thích
Implant loads are quarantined until the biological indicator reads negative, because a failure would place a permanently implanted device in a patient. Emergency release before the result is available is permitted only when it is documented and the surgeon is informed, and the result is tracked to completion. A Class 5 indicator supports early release decisions in an emergency but does not by itself satisfy the requirement, and modern rapid-readout indicators give results in far less than 48 hours.
Trích dẫn luật: AAMI ST79Luyện miễn phí toàn bộ 100 câu hỏi — không cần đăng ký.
Câu hỏi liên quan cùng chủ đề
- Which of the following are physical monitors of a sterilization cycle?
- A biological indicator from this morning's steam load is positive. What is the first action?
- After a positive biological indicator, the sterilizer is inspected, a probable cause is corrected, and testing resumes. What is generally required before the sterilizer returns to routine service?
- A recall requires the department to trace every package from a suspect cycle. What information must the load record contain?
- A nurse reports that a tray opened in the operating room contained an internal chemical indicator that did not change. What is the correct action?
- How should the daily Bowie-Dick test cycle be run?
Cập nhật gần nhất: · quy trình kiểm tra
Đội Ngũ Biên Tập PrepPass · Đối chiếu với HSPA Certified Registered Central Service Technician (CRCST) Exam · Quy trình kiểm tra