Monitoring & QACâu 72 / 100

How should the daily Bowie-Dick test cycle be run?

a.In a fully loaded chamber, to reflect real conditions
b.In an otherwise empty chamber, in the first cycle of the day after a warm-up cycle, at 270 to 275 degrees F for the time specified by the test manufacturer, usually not exceeding 4 minutes
c.In any cycle of the day, as long as the result is recorded
d.Together with a biological indicator, so both tests share the same cycle

Giải thích

The Bowie-Dick test is run in an empty chamber so the only challenge to air removal is the test pack itself, and a warm-up cycle first brings the chamber and jacket to operating condition. Exposure time follows the test manufacturer instructions and is typically limited to about 3.5 to 4 minutes, because overexposure can produce a false pass. Running it in a production load or combining it with biological monitoring invalidates the intent of the test.

Trích dẫn luật: AAMI ST79