Monitoring & QACâu 72 / 100
How should the daily Bowie-Dick test cycle be run?
a.In a fully loaded chamber, to reflect real conditions
b.In an otherwise empty chamber, in the first cycle of the day after a warm-up cycle, at 270 to 275 degrees F for the time specified by the test manufacturer, usually not exceeding 4 minutes
c.In any cycle of the day, as long as the result is recorded
d.Together with a biological indicator, so both tests share the same cycle
Giải thích
The Bowie-Dick test is run in an empty chamber so the only challenge to air removal is the test pack itself, and a warm-up cycle first brings the chamber and jacket to operating condition. Exposure time follows the test manufacturer instructions and is typically limited to about 3.5 to 4 minutes, because overexposure can produce a false pass. Running it in a production load or combining it with biological monitoring invalidates the intent of the test.
Trích dẫn luật: AAMI ST79Luyện miễn phí toàn bộ 100 câu hỏi — không cần đăng ký.
Câu hỏi liên quan cùng chủ đề
- An implant is processed in a terminal steam cycle with a rapid-readout biological indicator. When may the implant be released?
- A recall requires the department to trace every package from a suspect cycle. What information must the load record contain?
- A nurse reports that a tray opened in the operating room contained an internal chemical indicator that did not change. What is the correct action?
- What is the correct handling of the control biological indicator that accompanies a test lot?
- Which statement best describes the regulatory and standards landscape for sterile processing?
- A sterilizer has just been relocated to a new position in the department. What testing is required before it processes patient loads?
Cập nhật gần nhất: · quy trình kiểm tra
Đội Ngũ Biên Tập PrepPass · Đối chiếu với HSPA Certified Registered Central Service Technician (CRCST) Exam · Quy trình kiểm tra