What is the correct handling of the control biological indicator that accompanies a test lot?
a.It is placed in the sterilizer alongside the test indicator
b.It is discarded because it is not part of the record
c.It is not sterilized, is incubated with the processed indicator from the same lot, and must show growth for the test to be valid
d.It is stored in a refrigerator until the end of the month
Giải thích
The unprocessed control proves the spore population is viable and that the incubator and reading method are working, so it must turn positive while the processed indicator turns negative. If the control fails to grow, the test result is invalid regardless of what the processed indicator shows. Sterilizing the control defeats its purpose, and it must come from the same lot and be incubated at the same time.
Trích dẫn luật: AAMI ST79Luyện miễn phí toàn bộ 100 câu hỏi — không cần đăng ký.
Câu hỏi liên quan cùng chủ đề
- A recall requires the department to trace every package from a suspect cycle. What information must the load record contain?
- A nurse reports that a tray opened in the operating room contained an internal chemical indicator that did not change. What is the correct action?
- How should the daily Bowie-Dick test cycle be run?
- Which statement best describes the regulatory and standards landscape for sterile processing?
- A sterilizer has just been relocated to a new position in the department. What testing is required before it processes patient loads?
- A recall is initiated after a sterilization failure. Which action is part of a proper recall procedure?
Cập nhật gần nhất: · quy trình kiểm tra
Đội Ngũ Biên Tập PrepPass · Đối chiếu với HSPA Certified Registered Central Service Technician (CRCST) Exam · Quy trình kiểm tra