EKG AcquisitionPregunta 44 de 100
A tracing shows a uniform, thick, fuzzy baseline with small spikes occurring exactly 60 times per second across all leads. What is the most likely cause?
a.The patient is shivering
b.The electrodes are dried out
c.The patient is breathing deeply
d.Electrical interference from nearby powered equipment
Explicación
Sixty-cycle interference is alternating current picked up from nearby electrical devices, cords, or fluorescent lights, and it produces a perfectly regular fuzzy band at 60 hertz in the affected leads. Shivering produces irregular jagged somatic tremor rather than a uniform pattern. Deep breathing causes a slow wandering baseline, and dried electrodes produce erratic noise or a flat trace in one lead.
Practica las 100 preguntas gratis — sin registro.
Preguntas relacionadas de este tema
- Where is the V6 electrode placed?
- Where should the limb electrodes be placed for a standard 12-lead EKG?
- A 12-lead tracing shows an inverted P wave, inverted QRS, and inverted T wave in lead I, while the chest leads look normal. What is the most likely cause?
- An anxious, cold patient produces a tracing with irregular jagged spikes that obscure the small waveforms. What artifact is this and what is the best first correction?
- A tracing shows the baseline slowly drifting up and down in a wave pattern that matches the patient's breathing. What should the technician check first?
- What does the standardization mark on an EKG tracing indicate when it is 10 mm tall?
Última revisión: · proceso editorial
Equipo Editorial de PrepPass · Verificado con NHA Certified EKG Technician (CET) Exam · Cómo revisamos