SterilizationPregunta 31 de 100
What is a standard gravity displacement steam cycle for a wrapped instrument set at 250 degrees F (121 degrees C)?
a.3 minutes
b.10 minutes
c.30 minutes
d.60 minutes
Explicación
At the lower temperature of 250 degrees F, a gravity cycle needs about 30 minutes of exposure to achieve the required lethality, plus drying time. Shorter exposures of 3 or 10 minutes do not deliver sufficient lethality at that temperature for a wrapped set. Sixty minutes exceeds the validated requirement and unnecessarily stresses instruments; time and temperature are inversely related, so higher temperature permits shorter exposure.
Referencia Legal: AAMI ST79Practica las 100 preguntas gratis — sin registro.
Preguntas relacionadas de este tema
- What is the fundamental difference between a gravity displacement cycle and a dynamic-air-removal (prevacuum) cycle?
- A wrapped instrument set is being sterilized in a dynamic-air-removal steam sterilizer at 270 degrees F (132 degrees C). What is the typical minimum exposure time?
- A technician opens the department at the start of the day shift and must schedule the Bowie-Dick test for a dynamic-air-removal sterilizer. How often is the test required?
- A Bowie-Dick test sheet comes out of the sterilizer with a pale, uneven center and darker edges. What does this indicate and what should be done?
- After a tray fails its internal chemical indicator, the supervisor reviews the fundamentals with the team. Which four parameters must all be satisfied for steam sterilization to be effective?
- Why is saturated steam, rather than superheated or wet steam, required for steam sterilization?
Última revisión: · proceso editorial
Equipo Editorial de PrepPass · Verificado con HSPA Certified Registered Central Service Technician (CRCST) Exam · Cómo revisamos