SterilizationPregunta 52 de 100
A technician notices the sterilizer printout shows the chamber reached only 268 degrees F during a cycle set for 270 degrees F, although the chemical indicator changed color. What should be done?
a.Release the load, since the chemical indicator confirms sterility
b.Release the load but note the discrepancy in the log for the next preventive maintenance visit
c.Rerun the same load in the same sterilizer without further investigation
d.Do not release the load; the physical monitor shows the cycle parameters were not met, so the load is quarantined and the sterilizer is evaluated
Explicación
Physical monitors are an independent record of what the chamber actually did, and a load that did not meet its specified parameters cannot be released even if an indicator changed color. Chemical indicators respond to conditions but do not prove that all critical parameters were achieved. Rerunning without investigation ignores a possible equipment fault that will affect subsequent loads.
Referencia Legal: AAMI ST79Practica las 100 preguntas gratis — sin registro.
Preguntas relacionadas de este tema
- A department is deciding which sterilization method to use for a heat-stable stainless steel instrument set. Which method should be selected and why?
- A technician receives a jar of petroleum-based lubricant and a container of powder that must be sterilized. Which method applies, and what are representative parameters?
- Rigid sterilization containers are being loaded into a steam sterilizer along with wrapped trays. What is correct practice?
- Why must instruments with hinges, such as hemostats, be placed in the sterilizer in the open, unlocked position?
- A student asks a technician what it means when a process is described as achieving a sterility assurance level of 10 to the minus 6. What is the correct explanation?
- A technician is loading a steam sterilizer and wants to fit as many trays as possible before the shift ends. How full should the chamber be loaded?
Última revisión: · proceso editorial
Equipo Editorial de PrepPass · Verificado con HSPA Certified Registered Central Service Technician (CRCST) Exam · Cómo revisamos