SterilizationPregunta 51 de 100
Rigid sterilization containers are being loaded into a steam sterilizer along with wrapped trays. What is correct practice?
a.Stack containers three high to conserve shelf space
b.Place containers on the top shelf so condensate drains onto the wrapped trays below
c.Place containers flat on the shelf, do not stack unless the container manufacturer validates stacking, and keep them from touching chamber walls
d.Turn containers on edge so water drains from the filter ports
Explicación
Containers must sit level so condensate can drain through the designed pathway, and stacking is permitted only when the container manufacturer has validated it, because stacking can block steam access and trap moisture. Items should never contact chamber walls, where they can be scorched or collect condensate. Putting containers above absorbent wrapped packs invites wet packs, and turning them on edge defeats the drainage design.
Referencia Legal: AAMI ST79Practica las 100 preguntas gratis — sin registro.
Preguntas relacionadas de este tema
- An ozone sterilization system is being introduced. Which statement about ozone sterilization is accurate?
- A department is deciding which sterilization method to use for a heat-stable stainless steel instrument set. Which method should be selected and why?
- A technician receives a jar of petroleum-based lubricant and a container of powder that must be sterilized. Which method applies, and what are representative parameters?
- A technician notices the sterilizer printout shows the chamber reached only 268 degrees F during a cycle set for 270 degrees F, although the chemical indicator changed color. What should be done?
- Why must instruments with hinges, such as hemostats, be placed in the sterilizer in the open, unlocked position?
- A student asks a technician what it means when a process is described as achieving a sterility assurance level of 10 to the minus 6. What is the correct explanation?
Última revisión: · proceso editorial
Equipo Editorial de PrepPass · Verificado con HSPA Certified Registered Central Service Technician (CRCST) Exam · Cómo revisamos