Monitoring & QAPregunta 76 de 100
A recall is initiated after a sterilization failure. Which action is part of a proper recall procedure?
a.Retrieve only the items still on the sterile storage shelves
b.Notify the affected departments verbally and take no further action
c.Identify every item processed since the last negative biological indicator, retrieve and quarantine them, notify risk management and infection prevention if any were used on patients, and document the investigation and corrective action
d.Reprocess the affected items and release them without documenting the event
Explicación
An effective recall traces every package from the affected cycles using load records and lot control labels, pulls them from storage and from the procedural areas, and escalates when items have already been used so patient follow-up can be considered. The event, the root cause analysis, and the corrective actions are documented as part of the quality program. Limiting the search to shelf stock or skipping documentation defeats the purpose of traceability.
Referencia Legal: AAMI ST79Practica las 100 preguntas gratis — sin registro.
Preguntas relacionadas de este tema
- A recall requires the department to trace every package from a suspect cycle. What information must the load record contain?
- A nurse reports that a tray opened in the operating room contained an internal chemical indicator that did not change. What is the correct action?
- How should the daily Bowie-Dick test cycle be run?
- What is the correct handling of the control biological indicator that accompanies a test lot?
- Which statement best describes the regulatory and standards landscape for sterile processing?
- A sterilizer has just been relocated to a new position in the department. What testing is required before it processes patient loads?
Última revisión: · proceso editorial
Equipo Editorial de PrepPass · Verificado con HSPA Certified Registered Central Service Technician (CRCST) Exam · Cómo revisamos