第 4 章,共 7 章21% 占考试比重

Sterilization Process

Sterilization is the area where a small procedural shortcut produces an invisible failure. Sterility cannot be seen, smelled, or measured at the point of use, so the technician's job is to reproduce a validated process exactly and then to prove, with three independent kinds of evidence, that the process performed as validated. This is the third of three HSPA sections that tie at 21 percent each. The outline puts both halves of the job here: the methods, cycles, and loading on one side, and the quality tests, monitoring, failure response, and load documentation on the other. Learn the parameters cold, and learn the monitoring logic just as cold, because much of this material is tested by asking whether a parameter is correct or whether a release decision is defensible.

How Steam Sterilization Works

Steam sterilization kills microorganisms through moist heat, which coagulates and denatures the proteins the organism needs to live. The killing power does not come from the heat alone; it comes from condensation. When saturated steam meets a cooler surface it condenses back into water and releases its stored latent heat directly onto that surface, delivering a large, fast burst of energy exactly where it is needed. This single mechanism explains almost every rule in the chapter. Because sterilization depends on that steam-to-surface contact, four conditions must all be satisfied at once: enough time, the correct temperature, moisture in the form of saturated steam, and direct contact between steam and every surface of every device. If any one of the four fails, the load is not sterile no matter how well the other three performed. Residual soil, a closed box lock, trapped air, or an overpacked chamber all defeat the contact requirement even when time and temperature look perfect on the printout. A point that trips up many candidates: pressure is not one of the four parameters. Pressure exists only to raise the boiling point of water so steam can reach working temperatures of about 250 degrees F (121 degrees C) or 270 degrees F (132 degrees C). You never add pressure to sterilize; you add pressure to make hotter steam possible. Steam quality matters too. Ideal steam is roughly 97 percent dry with about 3 percent moisture. Steam that is too wet carries excess water and causes wet packs and poor penetration, while superheated steam behaves like hot dry air, loses the condensation mechanism, and would need dramatically longer exposure to sterilize. Steam is the preferred method for any device that tolerates heat and moisture, because it is nontoxic, inexpensive, rapid, and the most thoroughly validated process available; low-temperature methods exist only to fill the gap for devices steam would destroy. The goal of any sterilization process is stated as a sterility assurance level of 10 to the minus 6, meaning no more than a one-in-one-million probability that a single processed item is still nonsterile. That is a probability statement, not a count of organisms killed.

Four parameters must all be met: time, temperature, moisture, and direct contact
Any one of the four failing means the load is not sterile, regardless of the other three. Residual soil, a closed box lock, or trapped air all defeat the contact requirement.
AAMI ST79
Pressure is not a sterilization parameter
It exists only to raise water's boiling point so steam can reach 250 or 270 degrees F. An option listing pressure as a critical parameter is wrong.
AAMI ST79
Steam must be saturated, approximately 97 percent dry
Excess moisture produces wet packs and poor penetration; superheat destroys the condensation mechanism that does the killing.
AAMI ST79
Sterility assurance level is 10 to the minus 6
No more than a one-in-one-million probability that a single processed item remains nonsterile. It is a probability, not a body count.
AAMI ST79

Sterilizer Anatomy, Cycle Phases, and Daily Operator Care

HSPA expects you to know the machine, not just the numbers it produces. A steam sterilizer's chamber sits inside a jacket that is itself filled with steam. The jacket keeps the chamber walls hot between and during cycles, which reduces condensation on the walls and helps the load dry; it is a heat-management feature, not part of the sterilizing space. The door seals against a gasket, and a worn, cracked, or dirty gasket is one of the most common causes of a leak and an aborted cycle. At the bottom front of the chamber is the drain, fitted with a strainer and a thermostatic trap. Because air is heavier than steam it settles here, which makes this both the coldest point in the chamber and the place a temperature sensor and a process challenge device belong. The drain screen is cleaned daily, since a clogged drain traps condensate and produces wet packs. Prevacuum units add a vacuum pump, and every sterilizer has a printer or electronic recorder that produces the cycle record. Every cycle, regardless of method, runs through three functional phases. Conditioning removes air and brings the chamber to temperature, passively in a gravity cycle and by vacuum pulses or pressurized steam flushes in a dynamic-air-removal cycle. Exposure is the sterilizing phase, in which the chamber holds the set temperature for the set time, and this is the only phase whose time is the exposure time you memorize. Exhaust removes steam from the chamber and is followed on most cycles by a drying phase, often under vacuum, that pulls residual moisture out of the load. Knowing the phases makes the failure modes obvious: an air-removal problem is a conditioning failure, a temperature shortfall is an exposure failure, and a wet pack is usually an exhaust-and-dry failure. Daily operator maintenance keeps all of that working, and HSPA treats it as the technician's job rather than the engineer's. Per the sterilizer's instructions for use, the operator cleans the chamber drain screen and the chamber interior, inspects the door gasket for cuts, hardening, and debris, wipes the door seat, checks that the printer has paper and ink and produces a legible record, verifies that incubator temperature is in range and documented, and confirms the supplies the next shift will need are stocked. Sterilizers report faults as error codes, and the expectation is that the operator looks the code up in the manual, responds as the manual directs, takes the unit out of service if the process is in doubt, and documents and reports the fault. Clearing a code and rerunning the load is exactly wrong: an intermittent fault that is cleared and never reported is a fault that will eventually be discovered by a positive biological indicator instead.

The drain is the coldest point in the chamber
Air is heavier than steam and settles at the front bottom drain, which is why the temperature sensor and the process challenge device belong there. Clean the drain screen daily.
AAMI ST79
Every cycle runs conditioning, exposure, and exhaust with drying
Air removal problems are conditioning failures, temperature shortfalls are exposure failures, and wet packs are usually exhaust-and-dry failures.
AAMI ST79
The jacket manages heat; it is not part of the sterilizing space
It keeps chamber walls hot to limit condensation and aid drying. A worn or dirty door gasket is a leading cause of leaks and aborted cycles.
AAMI ST79
Look up an error code, do not clear and rerun it
Respond as the manual directs, take the unit out of service if the process is in doubt, and document and report the fault.
AAMI ST79

Gravity vs Dynamic-Air-Removal Cycles

Air is the enemy of steam sterilization, because any pocket of trapped air is a space steam never reaches. The two main steam cycle types differ entirely in how they remove it. In a gravity displacement cycle, incoming steam pushes the cooler, heavier air downward and out through a drain near the bottom of the chamber. This works, but it is passive and slow and struggles with porous loads, lumens, and dense sets where air can hide. In a dynamic-air-removal cycle, commonly called prevacuum, a vacuum pump actively pulls the air out in a series of pulses, or pressurized steam flushes drive it out, before steam is admitted. With the air already gone, steam contacts every surface almost immediately, so exposure times drop sharply. That difference shows up directly in the parameters. A wrapped set that needs about 15 minutes at 270 degrees F (132 degrees C) in a gravity cycle typically needs only about 4 minutes at the same temperature in a dynamic-air-removal cycle. At the lower temperature of 250 degrees F (121 degrees C), a gravity wrapped set needs roughly 30 minutes. Time and temperature are inversely related: raise the temperature and you may shorten the exposure. Drying time is added on top of exposure time and is often 30 minutes or more for wrapped goods. The overriding rule is that the device manufacturer's written instructions for use define the required cycle and minimum parameters for a specific instrument, and those instructions must be reconciled with what the facility's sterilizer can actually deliver before the device goes into service. Exposure parameters may never be shortened below the validated minimum for the sake of turnaround. And if a cycle aborts — the sterilizer cancels, alarms on an air leak, or the printout shows a parameter was not reached — the entire load is considered unsterile and reprocessed, regardless of what any chemical indicator shows.

Gravity wrapped set: about 30 minutes at 250 degrees F, or about 15 minutes at 270 degrees F
Passive air removal takes longer. Drying time is added on top of exposure and is commonly 30 minutes or more for wrapped goods.
AAMI ST79
Dynamic-air-removal wrapped set: about 4 minutes at 270 degrees F, or about 3 minutes at 275 degrees F
Active air evacuation lets steam reach surfaces immediately, which is why exposure is so much shorter than gravity at the same temperature.
AAMI ST79
The device instructions for use set the required minimum parameters
Parameters are never shortened below the validated minimum for turnaround. Conflicts between the device and the available sterilizer are resolved before the device goes into service.
AAMI ST79
An aborted cycle means the whole load is unsterile
If the sterilizer cancels, alarms on an air leak, or misses a parameter on the printout, the load is quarantined and reprocessed no matter what the indicators show.
AAMI ST79

Immediate-Use Steam Sterilization (IUSS)

IUSS is a steam process for items needed right away that will be transported directly to the point of use without storage. It is not a shortcut around cleaning. Every item processed by IUSS must first be fully decontaminated using the same cleaning steps as any other device, because a soiled instrument cannot be sterilized by any cycle; IUSS shortens only the packaging, drying, and storage steps. It should never be used for convenience, to make up for insufficient instrument inventory, or to save turnaround time, and the IUSS rate is a standing quality benchmark precisely because a rising rate signals an inventory or scheduling problem rather than a sterilization one. Parameters depend on the device. A common IUSS gravity cycle for nonporous metal instruments only is 270 degrees F (132 degrees C) for 3 minutes, while porous items, cannulated devices, and instruments with lumens typically require a longer exposure such as 10 minutes in a gravity cycle, always as specified by the device instructions for use. Implants deserve special attention. Because a failed implant sterilization has permanent consequences for the patient, implants should be terminally sterilized and quarantined until the biological indicator result is known. When a documented emergency makes IUSS of an implant unavoidable, a process challenge device containing a rapid-readout biological indicator plus a Class 5 integrating indicator is used, the surgeon is informed, the exception is documented as an early release, and the biological result is tracked to completion. Every IUSS cycle is still monitored: physical monitors are reviewed, a chemical indicator is used, and the sterilizer is biologically monitored on the schedule set by the standard and facility policy. IUSS items are used immediately and never stored, so there is no shelf life for an IUSS item.

IUSS never shortens cleaning
Full decontamination comes first. IUSS shortens only packaging, drying, and storage.
AAMI ST79
IUSS items are used immediately and never stored
The item is transported directly to the point of use in a validated containment device. There is no shelf life and no sterile storage for an IUSS item.
AAMI ST79
Common IUSS gravity parameters: 270 degrees F for 3 minutes, metal only
Porous, cannulated, and lumened items typically need a longer exposure such as 10 minutes in a gravity cycle, always per the device instructions.
AAMI ST79
Avoid IUSS for implants
If a documented emergency makes it unavoidable, use a process challenge device with a rapid-readout biological indicator and a Class 5 integrating indicator, notify the surgeon, and document the early release.
AAMI ST79

Low-Temperature Sterilization

Many modern devices contain plastics, optics, adhesives, or electronics that steam would destroy, so low-temperature methods exist to fill that gap, each trading away some of steam's convenience in exchange for a lower temperature. Ethylene oxide is a gas that sterilizes at low temperature and penetrates extremely well, which makes it the option for long, narrow lumens and complex heat-sensitive assemblies. Its trade-offs are severe: it is a recognized carcinogen, it requires strict engineering controls and monitoring under its own OSHA standard, and every load must undergo mechanical aeration to drive absorbed residuals out of plastics and rubber before items are safe to handle or use, commonly 8 to 12 hours or longer per the sterilizer and device instructions. Hydrogen peroxide gas plasma and vaporized hydrogen peroxide are much faster and leave no toxic residue, because the sterilant breaks down into water vapor and oxygen, so items are available immediately with no aeration phase. Their limitation is material compatibility: cellulose materials such as paper, linen, and cotton absorb hydrogen peroxide and will abort the cycle, so only nonwoven polypropylene wrap and approved trays and pouches may be used, and lumen restrictions apply. The sterilant arrives in cassettes or cartridges, which are handled with gloves per the instructions, never forced, and never handled at all if damaged or leaking, because a damaged cassette is a chemical spill managed with the safety data sheet and the spill protocol. Ozone sterilizers generate the sterilant on demand from medical-grade oxygen and water and revert it to oxygen at the end of the cycle, so there are no stored chemical cartridges and no toxic residue. Dry heat, a high-temperature method despite living next to these in most people's memory, remains the option for anhydrous materials such as oils, ointments, and powders that steam cannot penetrate, using long exposures such as about 320 degrees F (160 degrees C) for 2 hours or 340 degrees F (170 degrees C) for 1 hour. Liquid chemical sterilants such as peracetic acid process immersible devices in a chamber of circulating solution, but the item emerges wet and unwrapped, so there is no shelf life; it is transferred aseptically and used immediately. The unifying rule is to choose steam whenever the device tolerates heat and moisture, because every low-temperature method carries a cost in cycle time, material restriction, or worker safety.

Ethylene oxide requires mandatory aeration after every load
Residual EO absorbed into polymers is toxic and must be driven off, commonly 8 to 12 hours or longer at elevated temperature per the sterilizer and device instructions.
OSHA 29 CFR 1910.1047
No cellulose in hydrogen peroxide sterilizers
Paper, linen, cotton, and cellulose-based wraps absorb hydrogen peroxide and cause cycle cancellation. Use only compatible polypropylene wrap, pouches, and containers.
AAMI ST58
Hydrogen peroxide byproducts are water and oxygen
There is no toxic residual, so loads are used immediately with no aeration — the method's main operational advantage over ethylene oxide.
AAMI ST58
Dry heat is for anhydrous materials
Oils, ointments, and powders that steam cannot penetrate, at roughly 320 degrees F for 2 hours or 340 degrees F for 1 hour.
AAMI ST79
Liquid chemical sterilization produces no shelf life
The device comes out wet and unpackaged, so it is transferred aseptically and used at once. It can never be stored for a later case.
AAMI ST58
Choose steam whenever the device tolerates heat and moisture
Every low-temperature method carries cost, cycle time, material restriction, or worker safety burdens, so they are reserved for devices steam would destroy.
AAMI ST79

Loading, Drying, Cooling, and Wet Packs

How a load is arranged determines whether a technically correct cycle actually reaches the instruments. Before anything goes in, each package's integrity is checked one last time — no holes or tears, filters and locks intact, seals unbroken, external indicator present — because a compromised package will not be made sterile by the cycle. Packages need space so steam can circulate freely, and nothing should touch the chamber walls, where scorching and condensate collection occur; the chamber is loaded loosely, never packed tight, because overloading is a leading cause of both sterilization failure and wet packs. Metal and basin items go on lower shelves and textile packs on upper shelves, because condensate drips downward and absorbent textiles beneath metal is a classic wet-pack cause. Peel pouches are placed on edge in a rack with paper facing plastic so steam enters through the porous side and condensate can drain. Rigid containers sit flat and are not stacked unless the container manufacturer has validated stacking. Hinged instruments are sterilized open on stringers so steam reaches the mating surfaces of ratchets and box locks. Two more loading rules HSPA names: verify the sterilization method, confirming that every package in the load is validated for the cycle you are about to run, since a cellulose-wrapped tray in a peroxide chamber will abort the cycle and a low-temperature-only device in a steam chamber will be destroyed; and prioritize items sensibly, so a rapid-turnaround set does not sit while a routine load runs. Drying, cooling, and staging are part of the process rather than an afterthought. The load stays undisturbed on the cart in a low-traffic, draft-free area away from air vents until it is fully cool, commonly 30 minutes to an hour or more depending on load density, and cooling racks are placed away from doors, vents, and cold exterior walls. Packages are not touched, sorted, or stacked while warm. A package that is damp when handled is considered contaminated, because moisture creates a wicking path from the outer surface to the contents. A wet pack is never wiped, air-dried, or re-run through only a drying phase and then released; the contents are rewrapped and fully reprocessed, and the cause is investigated. Common wet-pack causes include overloading, dense packing, insufficient drying time, opening the door too early, clogged drains, faulty steam traps, wet incoming steam, and cooling on a cold surface or in a draft.

Metal below, textiles above, chamber loaded loosely
Condensate drips downward, so absorbent textile packs beneath metal trays reliably produce soaked wrappers. Nothing touches the chamber walls.
AAMI ST79
Peel pouches go on edge, paper to plastic
Stacking pouches flat or overlapping them traps air and moisture between layers and blocks steam from reaching the porous side.
AAMI ST79
Verify that every package in the load is validated for this cycle
A cellulose-wrapped tray in a peroxide chamber aborts the cycle; a low-temperature-only device in a steam chamber is destroyed.
AAMI ST79
A wet pack is a contaminated pack
Moisture wicks microorganisms from the outside of the package to the contents. Reprocess from the beginning; never air dry, wipe, or re-run only the drying phase.
AAMI ST79
Do not touch or move a load until it has cooled
Handling hot, damp packages drives contamination through the packaging. Cool on the cart, away from vents and traffic, for 30 minutes or longer.
AAMI ST79

Sterilizer Quality Tests and the Three Kinds of Monitoring

Sterility cannot be observed, so the department relies on three independent categories of monitor used together, never interchangeably. Physical monitors are the sterilizer's own instrumentation: gauges, digital displays, and the cycle printout or electronic record capturing time, temperature, and pressure throughout the cycle. They are the only monitor that reports what actually happened inside the chamber moment by moment, which makes them the first thing to check before releasing a load. The operator reviews the record for every cycle, confirms all specified parameters were reached and held for the required duration, and initials the record as a documented accountability step. If the printout shows a shortfall — a chamber that reached only 268 degrees F on a cycle set for 270 degrees F, for instance — the load is not released even if the chemical indicators changed color, because chemical indicators respond to conditions but cannot prove every critical variable was satisfied. A physical-monitor discrepancy is also a warning that the equipment may be developing a fault, so it triggers evaluation of the sterilizer rather than a simple rerun. Chemical indicators, covered by class in Chapter 3, show that a package experienced sterilizing conditions. Biological indicators are the only monitor that directly demonstrates lethality, because they contain a population of spores harder to kill than the microorganisms found on devices. Match the spore to the method: Geobacillus stearothermophilus for steam, hydrogen peroxide, and ozone; Bacillus atrophaeus for ethylene oxide and dry heat. Steam sterilizers are monitored biologically at least once each day of use, with every-load monitoring recommended because it dramatically narrows the recall population if a failure occurs, and every load containing an implant is monitored and quarantined until the result is negative. Every biological indicator test includes an unprocessed control from the same lot, incubated at the same time and temperature, which must show growth; if the control fails to grow the test is invalid no matter what the processed indicator shows, because the spores may have been nonviable or the incubator out of range, which is why incubator temperature is itself verified and documented. The indicator is not simply set on a shelf. It is placed inside a process challenge device, a test assembly presenting a challenge equal to or greater than the most difficult item routinely processed, positioned in the area hardest for the sterilant to reach — in a steam sterilizer typically the bottom shelf near the front, over the drain, where residual air settles. Selecting the right device, placing it correctly, and verifying its lot number are all part of the technician's job. Two equipment tests complete the picture. The vacuum leak test asks whether the chamber holds a vacuum, because a dynamic-air-removal sterilizer that leaks pulls room air back in during conditioning and defeats the very air removal the cycle depends on; it is run on the schedule the sterilizer's instructions set, commonly weekly, and after repairs. The Bowie-Dick test, which HSPA also calls the Dynamic Air Removal Test, asks whether air removal and steam penetration are working. It is run each day a dynamic-air-removal sterilizer is used, in an otherwise empty chamber, in the first cycle after a warm-up cycle, at 270 to 275 degrees F for the time specified by the test manufacturer, usually not exceeding about 4 minutes because overexposure can produce a false pass. It tests air removal, not lethality, and a pale, blotchy, or uneven color change means air removal failed and the sterilizer comes out of patient service until the cause is corrected. Equipment testing is triggered by events as well as by the calendar: after installation, after relocation, after construction or utility work affecting the unit, and after any major repair or malfunction, the sterilizer is requalified before it processes patient loads.

Review and initial the physical record for every cycle before releasing the load
Time, temperature, and pressure are verified against the cycle specification. A shortfall quarantines the load and triggers evaluation of the sterilizer, not a rerun.
AAMI ST79
Physical, chemical, and biological monitors are used together, never interchangeably
Each answers a different question, and no single monitor of any type establishes sterility on its own.
AAMI ST79
Match the spore to the method
Geobacillus stearothermophilus for steam, hydrogen peroxide, and ozone; Bacillus atrophaeus for ethylene oxide and dry heat.
AAMI ST79
Steam biological indicator at least daily, every load preferred, every implant load required
Implant loads use a process challenge device with a biological indicator and are quarantined until the result is negative, except in a documented emergency.
AAMI ST79
The unprocessed control must grow
Same lot, never sterilized, incubated alongside the test indicator. No growth in the control invalidates the test. Incubator temperature is verified and documented.
AAMI ST79
Place the process challenge device in the coldest, most challenging location
In a steam sterilizer, generally the bottom shelf near the front over the drain, where residual air settles. Follow the sterilizer manufacturer instructions.
AAMI ST79
Bowie-Dick daily, empty chamber, first cycle after warm-up
It tests air removal and steam penetration, not lethality, at 270 to 275 degrees F for the time the test manufacturer specifies, usually no more than about 4 minutes because overexposure can mask a failure.
AAMI ST79
Run the vacuum leak test on the sterilizer's schedule, commonly weekly, and after repairs
A leaking dynamic-air-removal chamber pulls room air back in during conditioning and defeats air removal entirely.
AAMI ST79

Process Failures, Recall, and Load Documentation

Traceability is what turns a sterilization failure from a department-wide crisis into a targeted retrieval. Every load record documents the sterilizer identification, cycle or load number, date and time, contents, cycle parameters achieved, operator identification, and the results of physical, chemical, and biological monitoring. Every package carries a lot control label with the sterilizer, cycle number, and date, so it can be matched back to a specific record and any load can be traced forward to its packages. Records may be kept manually or in a tracking system, and their purpose is exactly three things: recall, traceability, and retention as a legal record. Process failures come in recognizable forms — a positive biological indicator, a wet pack, an unchanged internal chemical indicator, an aborted cycle, and a printout showing a parameter was not met — and the response to any of them starts the same way: do not release the load. When a biological indicator is positive, the sterilizer is taken out of service and every item processed since the last negative biological indicator is identified, retrieved, and quarantined. Positives are investigated, not dismissed as false. Items already distributed must be located rather than released for use, and if any were used on patients, risk management and infection prevention are notified so patient follow-up can be considered. Continuing to process while waiting to retest is exactly the wrong move, because it widens the population of potentially nonsterile items reaching patients. After a probable cause is identified and corrected, the sterilizer is requalified with three consecutive negative biological indicator cycles in an empty chamber, plus three passing Bowie-Dick tests for dynamic-air-removal units, before returning to routine service, and the same qualification testing is performed after installation, relocation, and major repairs. The entire event, its root cause analysis, and the corrective action are documented in the quality program.

Every package carries a lot control number
Sterilizer identification, cycle number, and date, so any package traces back to its load record and any load traces forward to its packages.
AAMI ST79
A positive biological indicator triggers immediate recall back to the last negative
Take the sterilizer out of service, retrieve and quarantine every affected load, and never keep processing while waiting for a retest.
AAMI ST79
Requalification is three consecutive empty-chamber cycles with negative biological indicators
Dynamic-air-removal sterilizers add three passing Bowie-Dick tests. Required after a failure and also after installation, relocation, or major repair.
AAMI ST79
Escalate when affected items reached patients
Notify risk management and infection prevention, and document the investigation and corrective action in the quality record.
AAMI ST79
Load documentation exists for recall, traceability, and legal defense
Sterilizer, cycle number, date and time, contents, parameters achieved, operator, and all monitoring results, kept manually or in a tracking system.
AAMI ST79
测试你的知识
练习 Sterilization Process 的相关题目
立即练习 →

Last updated: September 2026

反馈